HeartPlus NT-180 is a compact, portable device that analyses the heart’s rhythm and delivers a defibrillation shock to re-establish the normal heart rhythm. HeartPlus NT -180 guides the user through each step of the defibrillation process by providing voice/ visual prompts.
HeartPlus NT -180 of Nanoomtech Co., Ltd provides Uni-Pads which can be used for both adults and child and two types of cartridges, integrated-type or separated-type.

TECHNICAL DETAILS:
Energy : 180J(Adult), 50J(Child)
Shock Charge Time : Approx. 13 Seconds
Lead 2
Analysis Time : Less than 11 Secons
ECG 1,000 instance of 5 seconds recordings
Ambient Noise : Total 60 Minutes (4 instances of 15 minute cycles)
Self-Test: Times & Result of 3,000 events
Product Highlights
Defibrillator

Defibrillator Series HeartPlus NT-180
Set Configuration Defibrillator, battery pack, sealed electrode pads, HeartPlus User Guide, Protective Carrying Case, “NT-MPR” CD-ROM USB cable
Waveform Truncated Exponential Biphasic
Energy 180J (adult, impedance 50?) / 50J (pediatric, impedance 50?)
ShockChargelimeI Less than 13 seconds
Shock Method Manual button (flashes when ready)
Mode Conversion Adult/Child separate power buttons Child Mode delivers 50J Pediatric shock for patients weighing less than 25kg, approximately 12 months to 8 years old)
Physical

Dimensions 7.1cm (H) x29.3cm (W) x29.1cm (D)
Weight 1.9kg (including battery pack and electrode pad)
Environmenta

Water resistance IPX2 based on IEC 60529
Temperature Operating : 0? to 40? / Storage : -20? to 60?
Humidity
(operating/storage) 10% to 95% Relative Humidity, Non-Condensing
Altitude
(operating/storage) 0 to 2000 meters
Drop test 1M
ECG

ECG Record LEAD II
Analysis Overview Patient evaluated to determine whether ECG rhythm represents shockable or non-shockable state. Shockable rhythms are ventricular fibrillation (VF) and ventricular tachycardia (VT).
Analysis time Less than 11 seconds (Read: 5 seconds ± 2 seconds, Analyze: within 4 seconds)
Audio/ Visual instruction

After device is powered on, step-by-step voice instructions guide first responder through resuscitation procedure, which includes pads placement, hands offtimes (analysis and shock process), and CPR
CPR instruction prompts user to commence cardiopulmonary resuscitation, while flashing metronome light sets pace at 100 compressions per minute for 30 compressions with brief pause for 2 mouth-to-mouth breaths (5 repetitions), in accordance with American Heart Association 2010 guidelines.
Data

Storeda data ECG recordings, operation information (whether shock delivered or not), ambient noise, self-diagnostics history
Memory capacity ECG: 1,000 instances of 5 seconds recordings Ambient
Noise: Total 60 minutes (4 instances of 15 minute cycles)
Self-Test: Times & results of 3,000 events
Storage type Internal Memory
Access/search “NT-MPR” App (for PC) via USB connection
self-diagnostics Test

Daily self-diagnostics tests ensure that circuits for measuring patient impedance, control circuit internal discharge of electric energy, and control circuit for charge/discharge of electric energy are functioning
A voice prompt stating “Equipment check is needed” continuously in the event self-diagnostics check fails.
Upon completion of daily self-diagnostics test, results are automatically stored within device’s internal memory.
Battery Check

When battery reserve is at acceptable level to provide sufficient electric charges and there are no problems with circuitry detected during self check, there Status LEDs above the power button blink in order from left to right during self check.
All three Status LEDs blink at the same time continuously with low or critical battery voice prompts when battery is at low or critical levels (low: shock possible, critical: shock not possible).
Battery Status Audible Warnings Low: Voice prompt states “Battery’s voltage is low, check battery”
Critical: Voice prompt states “Change cartridge”
Electrode pads

Shelf Life Two (2) years, for one-time use only
Conductive Gel PE foam, Hydrogel
Conductive Element AgCl (Silver Chloride)
Cable Length 2.1m (+1%)
Applicable Ages All ages (12 months & older)
Sealing Airtight
Size 125mm x 95mm
Pads Placement Anterior-lateral (visual details on pads)
Battery Cartridge

Type Non-rechargeable lithium (LiMn02)
Capacity 180 shocks @ 25? ,130 shocks @ 0?
Voltage DC 21V& DC V / 1,400mAH
Standby Mode Life Recommended five (5) years @ 25?
Dimensions 3.3cm (H) x 15.0cm (W) x 11.0cm (D)

Original price was: 4,600 AED.Current price is: 4,150 AED.

17 People watching this product now!

Additional information

Customer Reviews

Reviews

There are no reviews yet.

Be the first to review “Heart Plus AED (Automated External Defibrillator)”

Your email address will not be published. Required fields are marked *

You have to be logged in to be able to add photos to your review.